Work Item
ASTM WK84104
New Test Method for Determining the Particle and Inclusion size, and count in Nitinol products
1. Scope
This test method is used to determine the particle or inclusion size and the count per unit area (number of particles per mm2) using light microscopical images. Images generated by other devices, such as a scanning electron microscope, are not specifically addressed, but can be utilized if the system is calibrated in both x and y directions.Keywords
Nitinol; Image Analysis; Second phase particles in nitinolRationale
Material Specification for Nitinol products specify different levels of purity or freedom from second phase particles and inclusions however there is no standard test method to evaluate the level of purity or freedom from second phase particles and inclusions. This test method outlines a repeatable standard process to use image analysis for such and evaluation. Both producers and users can use this test method for evaluating product.
Work Item Status
Date Initiated: 11-05-2022
Technical Contact: Dr. Matthias Frotscher
Item: 001
Ballot: E04.14 (24-01)
Status: Will Reballot Item
Item: 001
Ballot: E04 (24-03)
Status: Negative Votes Need Resolution