Work Item
ASTM WK84901
New Test Method for Particulate and Coating Integrity Characterization of Coated Vascular Stents and Coated Endovascular Prostheses Subjected to Chronic Cyclic Loading
1. Scope
This standard provides guidance regarding in vitro conditions, particulate sizing, counting and identification, coating integrity inspections, test equipment, method validation, and reporting requirements for particulate and coating integrity characterization of coated vascular stents and endovascular devices subjected to chronic, cyclic loading conditions.Keywords
particle; accelerated; deployment; SEM; EDX; microscope; magnification; recovery; cyclic; expansion; deflection; high speed photography; camera; temperature; particle counter; DES; baseline; coating inspection; drug eluting vascular stent; absorbableRationale
Coating of stents is common and the long-term durability of these coatings needs to be evaluated. USP 788 is not appropriate for this evaluation, nor is there another standard that is applicable. It will be used to provide guidance on how to fulfill regulatory testing requirements aimed at the safety and efficacy of these products for the benefit of patients. The users will be manufacturers of coated stents, test labs and regulatory agencies.
Work Item Status
Date Initiated: 01-20-2023
Technical Contact: Kendra Conti
Item: 001
Ballot: F04.30 (23-01)
Status: Will Reballot Item
Item: 021
Ballot: F04 (24-04)
Status: Negative Votes Need Resolution