Work Item
ASTM WK86470
Revision of E2546-15 Standard Practice for Instrumented Indentation Testing
Rationale
Revision Request #1: Make A1.3.4 (timing verification) optional. Justification: Timing verification is onerous and not necessary for modern computer-controlled equipment. Also, ASTM E92 (incorporated by reference) makes timing verification optional. Revision Request #2: Add text to Section A1 and Appendix X4 that also allows indirect verification by HV using test blocks made in accordance with ASTM E92.A4. Justification: Verification by HV would be persuasive to the user and might be easier for some manufactures. Jennifer Hay is willing to draft text for these changes.
Work Item Status
Date Initiated: 05-18-2023
Technical Contact: Jennifer Hay
Item: 000
Ballot:
Status: