ASTM WK93004

Revision of D6122-23 Standard Practice for Validation of the Performance of Multivariate Online, At-Line, Field and Laboratory Infrared Spectrophotometer, and Raman Spectrometer Based Analyzer Systems

Rationale

Two issues 1) Control chart – failure criteria, 1 in 20 failures, needs to be reviewed reviewed. Solution a) include EWMA or MCEWMA from D6299 to define failure of control chart. Solution b) use rolling windows for n value for sample size. 2) Total number of samples required to move to general validation. Right now it recommend K samples but needs further clarification

Details

Developed by Subcommittee: D02.25

Committee: D02

Staff Manager: Alyson Fick

Work Item Status

Date Initiated: 11-25-2024

Technical Contact: Bryan Bowie

Item: 000

Ballot: 

Status: 